Article number: 5766

SYMPTOMS

  • The mass spectrometer is not communicating with the acquisition computer
  • Waters DHCP server does not assign an IP address to the MS, in MassLynx or UNIFI
  • EPC fails to boot, but the UPLC is communicating properly

ENVIRONMENT

  • UNIFI
  • Windows 7
  • MassLynx
  • 3100 Mass Detector
  • SQ Detector
  • SQ Detector 2
  • TQ Detector
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Vion IMS QTof
  • Xevo QTof
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

The Windows firewall is enabled for the instrument LAN card.

FIX or WORKAROUND

  1. Close the acquisition software.
  2. Disable the Windows firewall for the instrument LAN card (Guest or public networks).
  3. Reboot the embedded PC of the mass spectrometer.
  4. Launch the acquisition software to verify that the mass spectrometer and the acquisition PC communicate.

ADDITIONAL INFORMATION

 

id5766, A-SQ, A-SQ, comms, MALDIG2SI, MLYNX, MLYNXV41, SQD, SQD2, SUPMM, SUPUNIFI, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVTQSMIVD