Why doesn't the PEEK handle fit my new cone gas nozzle? - WKB5740
Article number: 5740
ENVIRONMENT
- SQD2
- SYNAPT G2
- SYNAPT G2-S/Si
- Xevo G2 Tof
- Xevo G2 QTof
- Xevo G2-S
- Xevo G2-XS
- Xevo QTof
- Xevo TQ
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-Sµ
ANSWER
A recent change in manufacturing process means that the cone gas nozzle no longer has a screw thread for attaching the PEEK handle.
The PEEK handle was redesigned to push-fit over the barbed arm.
ADDITIONAL INFORMATION
The part number for the cone gas nozzle has not changed; it must be ordered as an assembly (700004216) because the handle must be fitted in-house using a specialized tool.
The new handle is not removed for cleaning. PEEK is resistant to a wide range of solvents at 60 ºC or lower, so the cone gas nozzle can be cleaned as an assembly.