No spray seen from electrospray probe when infusing at a flow rate between 10 and 100 µL/min - WKB57255
Article number: 57255
SYMPTOMS
- No spray seen when infusing at flow rate between 10 and 100 µL/min
- Spray looks fine when infusing at 200 µL/min
- Spray looks fine when flowing from UPLC
ENVIRONMENT
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ-S micro
CAUSE
Partial blockage in the electrospray probe capillary
FIX or WORKAROUND
- Flush strong solvent through the electrospray probe for several hours to remove any compounds that may have precipitated.
- Flush aqueous solution through the electrospray probe for several hours to flush out any salts that may have precipitated.
- Repeat steps 1 and 2 as needed until spray is stable when infusing.
- If flushing does not unblock the probe, then rebuild it using part 700004672.
ADDITIONAL INFORMATION
id57255, ESI, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD