Low sensitivity with a method on a TQ-S micro - WKB57401

Article number: 57401

SYMPTOMS

  • Lower than expected sensitivity when comparing results with those from a second instrument
  • Cleaning the source ion optics does not improve sensitivity

ENVIRONMENT

  • Xevo TQ-S micro

CAUSE

Wrong MRM transition specified in the MS method

FIX or WORKAROUND

Create an MS method that uses the correct transition for the analyte.

ADDITIONAL INFORMATION

 

id57401, XEVOTQSMIC