Low MRM sensitivity using the Xevo TQ-S micro - WKB57401
Article number: 57401
SYMPTOMS
- Lower than expected MRM sensitivity when comparing results with those from a second instrument
- Cleaning the source ion optics does not improve sensitivity
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Wrong MRM transition specified in the MS method
FIX or WORKAROUND
Create an MS method that uses the correct transition for the analyte.
ADDITIONAL INFORMATION