No flow from fluidics on Xevo/SYNAPT/Vion - WKB50419

Article number: 50419

SYMPTOMS

  • No beam from infusion
  • With PEEK disconnected from source, no liquid flow is seen when trying to infuse from bottles
  • Flow can be seen when infusing wash solution (after purging)

ENVIRONMENT

  • Synapt HDMS
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • Xevo G3
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ XS

CAUSE

The orange PEEK tubing does not reach down to the bottom of the 30-mL Nalgene fluidics bottle; therefore, air is drawn instead of sample.        

FIX or WORKAROUND

  1. Open the fluidics door.
  2. Loosen the finger-tight fitting (PEEK on the left is for reference syringe; right is for sample syringe).
  3. Pull some tubing through above the finger-tight from the instrument interior.
  4. Feed it down into the bottom of the 30-mL bottle.
  5. Tighten the finger-tight to lock in place.
  6. Close the fluidics door.
  7. Purge fluidics several times if needed, and check flow.

ADDITIONAL INFORMATION

 

id50419, MALDIG2SI, SYNG2SIHD, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOG3QTOF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD