No flow from fluidics on Xevo/SYNAPT/Vion - WKB50419

Article number: 50419

SYMPTOMS

  • No beam from infusion
  • With PEEK disconnected from source, no liquid flow is seen when trying to infuse from bottles
  • Flow can be seen when infusing wash solution (after purging)

ENVIRONMENT

  • Synapt HDMS
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ XS

CAUSE

The orange PEEK tubing does not reach down to the bottom of the 30-mL Nalgene fluidics bottle; therefore, air is drawn instead of sample.        

FIX or WORKAROUND

  1. Open the fluidics door.
  2. Loosen the finger-tight fitting (PEEK on the left is for reference syringe; right is for sample syringe).
  3. Pull some tubing through above the finger-tight from the instrument interior.
  4. Feed it down into the bottom of the 30-mL bottle.
  5. Tighten the finger-tight to lock in place.
  6. Close the fluidics door.
  7. Purge fluidics several times if needed, and check flow.

ADDITIONAL INFORMATION

 