"Leak detected" error with TQ Detector - WKB50404
Article number: 50404
SYMPTOMS
- "Leak detected" error in the console
ENVIRONMENT
- TQ Detector
CAUSE
Blocked PEEK tubing from the diverter valve to the probe
FIX or WORKAROUND
Replace the PEEK tubing.
ADDITIONAL INFORMATION
id50404, TQD, TQDIVD