Skip to main content
Waters

Fault in the waste trap causes source pressure test fails - WKB49439

Article number: 49439

SYMPTOMS

  • The source pressure test fails
  • The system reaches +200 mBar shortly after the test starts
  • An exhaust problem warning is triggered
  • The source pressure test fails after opening and closing the source enclosure
  • A new source enclosure O-ring does not resolve the issue

ENVIRONMENT

  • Xevo TQ-MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof

CAUSE

A fault in the waste trap bottle assembly caused the exhaust solenoid valve to permanently shut, causing a blockage in the source exhaust system and resulting in the high pressures observed with gases on.

FIX or WORKAROUND

Replace the Bottle Waste Trap Assembly (700004606).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.