What is the part number of the sample cone for TQ and QTof instruments? - WKB4678
ENVIRONMENT
- 3100 Mass Detector
- SELECT SERIES Cyclic IMS
- SELECT SERIES MRT
- SQ Detector
- SQ Detector 2
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Vion IMS QTof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S QTof
- Xevo G2-XS QTof
- Xevo G3 Qtof
- Xevo MRT
- Xevo QTof
- Xevo TQ MS
- Xevo TQ-Absolute
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
ANSWER
This is to notify the different size of cone related to Waters Mass Spec instruments. They are NOT interchangeable between the instruments. You should ONLY use the cone appropriate for the instrument it is intended for. Please see the attached photos to see where the part numbers are located.
cone gas nozzle |
sample cone o-ring |
Sample cone |
aperture |
Instrument |
|
700004216 |
700004321 |
0.25 mm |
Xevo TQ-S cronos |
|
|
700004216 |
700004321 |
0.35 mm |
Xevo QTof |
|
|
700004216 |
700004321 |
0.45 mm |
Xevo TQ-S micro |
|
|
700004216 |
700004321 |
0.5 mm |
3100 Mass Detector |
|
|
700004216 |
700004321 |
0.5 mm |
SQ Detector 2 |
|
|
700004216 |
700004321 |
0.67 mm |
Xevo G2-S QTof |
|
|
700004216 |
700004321 |
0.8 mm |
SYNAPT G2-Si |
|
|
700004216 |
700004321 |
0.9 mm |
Xevo TQ-XS |
ADDITIONAL INFORMATION
Inspect the sample cone o-ring for damage or replace with a new one if needed.
Sample cone o-ring (# 700004321) size 10 mm x 1.5 mm
Cone gas nozzle (# 700004216)
Handle, Cone Gas Nozzle (# 700010292)
Cone gas o-ring ( # 700004728) size 009 (o-ring between sample cone and cone gas nozzle)