The argon tank is emptying in a quarter of its normal time - WKB46737
Article number: 46737
SYMPTOMS
- Argon tank empties too quickly
ENVIRONMENT
- Xevo TQ MS
- Xevo G2-XS
CAUSE
The Swagelok connection is too loose.
FIX or WORKAROUND
Tighten the 1/8-inch Swagelok connection.
ADDITIONAL INFORMATION
id46737, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQIVD