Desolvation temperature does not reach set point on the Quattro Micro, SQD2, Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S Micro or Xevo TQ-XS - WKB46567
Article number: 46567
SYMPTOMS
- Desolvation temperature does not reach set point
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- SQD2
- Quattro micro API
CAUSE
Unknown board error
FIX or WORKAROUND
Reboot the EPC.
ADDITIONAL INFORMATION
id46567, A-SQ, QUMICRO, SQD2, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD