Cannot pass sensitivity specification on a Xevo TQ-S or SQD2 - WKB45658
Article number: 45658
SYMPTOMS
- %RSD height is outside of tolerance for sensitivity specification
ENVIRONMENT
- Standalone MS and qualification test
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- SQD2
CAUSE
Unstable fluidics
FIX or WORKAROUND
During the run, do a purge after each sample injection.
ADDITIONAL INFORMATION
id45658, A-SQ, SQD2, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD