Source pressure test failure due to leak - WKB3892
Article number: 3892
SYMPTOMS
- Source pressure fails to reach 100 mbar
- Dashboard trace shows correct gas supply but indicates small leak
- Error "Source pressure test leak warning" in console
- Error "Source pressure test failed" in console
- Vacuum levels are within expected ranges
- Waste bottle and exhaust connections are secure
- Hinge fixing points are secure
- Catch dowel of source is undamaged
ENVIRONMENT
- Xevo TQ MS
- Xevo G2-XS
CAUSE
Damaged O-ring causes a minor front end vacuum leak, which prevents the source from pressurizing to 100 mbar.
FIX or WORKAROUND
Replace the BS039 O-ring (5711130) between the ion block support and the pumping block subassembly.
ADDITIONAL INFORMATION
Also see source pressure test advice note.
id3892, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQIVD