Source pressure test failure due to leak - WKB3892

Article number: 3892

SYMPTOMS

  • Source pressure fails to reach 100 mbar
  • Dashboard trace shows correct gas supply but indicates small leak
  • Error "Source pressure test leak warning" in console
  • Error "Source pressure test failed" in console
  • Vacuum levels are within expected ranges
  • Waste bottle and exhaust connections are secure
  • Hinge fixing points are secure
  • Catch dowel of source is undamaged

ENVIRONMENT

  • Xevo TQ MS
  • Xevo G2-XS

CAUSE

Damaged O-ring causes a minor front end vacuum leak, which prevents the source from pressurizing to 100 mbar.

FIX or WORKAROUND

Replace the BS039 O-ring (5711130) between the ion block support and the pumping block subassembly.

ADDITIONAL INFORMATION

Also see source pressure test advice note.

id3892, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQIVD