Flat baseline in MRM acquisitions on Xevo TQ-S - WKB3395
Article number: 3395
SYMPTOMS
- Flat baseline in MRMs
- No peaks seen where expected
ENVIRONMENT
- Xevo TQ-S
- MassLynx
CAUSE
Dwell time was too small for the method, causing a loss of signal.
FIX or WORKAROUND
- Adjust the dwell time to an appropriate value.
ADDITIONAL INFORMATION
id3395, MLYNX, MLYNXV41, SUPMM, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD