Flat baseline in MRM acquisitions on Xevo TQ-S - WKB3395

Article number: 3395

SYMPTOMS

  • Flat baseline in MRMs
  • No peaks seen where expected

ENVIRONMENT

  • Xevo TQ-S
  • MassLynx

CAUSE

Dwell time was too small for the method, causing a loss of signal.

FIX or WORKAROUND

  1. Adjust the dwell time to an appropriate value.

ADDITIONAL INFORMATION