Mass spectrometer status for QDa in MassLynx remains "Waiting for inlet start" after injection is made - WKB3404
Article number: 3404
SYMPTOMS
- System was recently moved
- QDa does not start acquisition
- QDa Status remains "Waiting for Inlet Start 1" after injection
- Real-time chromatogram display does not open
- Message at end of run: "No data collected"
ENVIRONMENT
- Fraction Collection System
- 2767 Sample Manager
- ACQUITY UPLC PDA Detector
CAUSE
A broken contact closure wire
FIX or WORKAROUND
- Connect a new/repaired contact closure wire between pins 1 and 2 on the 2767 and pins 5 and 6 on the QDa.
- Connect a new/repaired contact closure wire between pins 1 and 2 on the 2767 and pins 1 and 2 on the PDA.
ADDITIONAL INFORMATION
The two contact closure wires are piggy backed in the terminal connector, pins 1 and 2, on the 2767 Sample Manager.
id3404, 2767, 2767SFC, FLYNXV41, QDA, UNIFSW17