Intermittent beam when infusing samples on Quattro Premier XE - WKB32416
Article number: 32416
SYMPTOMS
- Intermittent beam when directly infusing samples
- Direct infusion sample carryover when switching between samples
ENVIRONMENT
- Quattro Premier XE
- MaxEnt
CAUSE
Void between top of ESI probe union and top of capillary
FIX or WORKAROUND
Rebuild the ESI probe, replacing the capillary, liner, and ferrule.
ADDITIONAL INFORMATION
id32416, Electrospray, QUATPREMXE, QUPREMIER