Skip to main content
Waters

Intermittent beam when infusing samples on Quattro Premier XE - WKB32416

Article number: 32416

SYMPTOMS

  • Intermittent beam when directly infusing samples
  • Direct infusion sample carryover when switching between samples

ENVIRONMENT

  • Quattro Premier XE
  • MaxEnt

CAUSE

Void between top of ESI probe union and top of capillary

FIX or WORKAROUND

Rebuild the ESI probe, replacing the capillary, liner, and ferrule.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.