No signal during direct infusion - WKB31657
Article number: 31657
SYMPTOMS
- Little or no signal at all during infusion
- Isolation valve is open
- Little or no spray from capillary
ENVIRONMENT
- Xevo QTof
- Xevo G2-XS
- Xevo G2-S
- Xevo G2
- Xevo TQ-S
- Xevo TQD
- Xevo TQ MS
- XevoTQ-S micro
- Xevo TQ-XS
- Vion
- SYNAPT G2-Si
CAUSE
The syringe pump was not purged correctly.
FIX or WORKAROUND
- Verify that the wash PEEK tube is completely submerged in solvent.
- Purge the syringe pump from the Tune page: Setup > Wash fluidics at least 10 times.
- Prime each reservoir 5 times with isopropyl alcohol.
- Verify that the wash solvent flows up from the waste line during purge.
- Verify that the signal is present during infusion.