Spray from Quattro LC probe is pulsing - WKB31626

Article number: 31626

SYMPTOMS

  • Spray is pulsing
  • Probe recently rebuilt

ENVIRONMENT

  • Quattro Premier
  • Quattro Premier XE
  • Quattro LC
  • Quattro LCZ
  • Quattro micro API
  • Quattro Ultima
  • 3100 Mass Detector
  • ACQUITY SQD system
  • SQ Detector
  • ACQUITY TQD system
  • TQ Detector
  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • ZQ 2000
  • ZQ 4000
  • ZMD

CAUSE

The capillary is partially blocked.

FIX or WORKAROUND

Replace the capillary (700000341 TUBE, ESI-SS CAPILLARY, PLCZ), or the capillary specific to the MS model.

ADDITIONAL INFORMATION

id31626, Electrospray, QUATTROLCZ