Spray from Quattro LC probe is pulsing - WKB31626
Article number: 31626
SYMPTOMS
- Spray is pulsing
 - Probe recently rebuilt
 
ENVIRONMENT
- Quattro Premier
 - Quattro Premier XE
 - Quattro LC
 - Quattro LCZ
 - Quattro micro API
 - Quattro Ultima
 - 3100 Mass Detector
 - ACQUITY SQD system
 - SQ Detector
 - ACQUITY TQD system
 - TQ Detector
 - SQ Detector 2
 - Xevo TQ MS
 - Xevo TQD
 - Xevo TQ-S
 - Xevo TQ-S cronos
 - Xevo TQ-S micro
 - Xevo TQ-XS
 - ZQ 2000
 - ZQ 4000
 - ZMD
 
CAUSE
The capillary is partially blocked.
FIX or WORKAROUND
Replace the capillary (700000341 TUBE, ESI-SS CAPILLARY, PLCZ), or the capillary specific to the MS model.
ADDITIONAL INFORMATION
id31626, A-SQ, A-SQ, Electrospray, EMD1000, QUATPREMXE, QUATTROLCZ, QUMICRO, QUPREMIER, SQD, SQD2, TQD, TQDIVD, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD, ZMD2, ZMD4, ZQ2000, ZQ2000SFC, ZQ4000
 