Intermittent communications dropout or no connection to Xevo TQ-XS or Xevo TQ-S - WKB30889
Article number: 30889
SYMPTOMS
- Multiple possible symptoms:
- No communications with MS but can connect to inlet
- Console message: MS communication failure.
- Intermittent communications
- Sample list pausing randomly
- System is not able to perform any MS/MS, and then communication is lost
- MassLynx displays the error message “Application Lancaster has encountered a problem and needs to close”
- Rebooting the host PC does not fix the problem
- Hard rebooting of the MS (this sometimes needs to be done several times) temporarily fixes the fault
- No communication between instrument and PC
- Might be able to Ping EPC
- Hyperterminal will not boot
- Or errors in Hyperterminal log such as:
-
Failed to save file to flash.
Get failed!
Failed to sync file, trying local copy...
get_file OK.
3445504Error reading ELF32 load segment.
Load Module failedError loading file: errno = 0x610007.
Can't load boot file!! - MassLynx Microsoft Visual C++ Runtime Library Runtime error!
R6025 - pure virtual function call.
-
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQ-S
CAUSE
Faulty EPC
FIX or WORKAROUND
- Replace EPC.
- Import treg file.
- Reboot via hyperterminal.
Xevo TQ-XS = (700013576)
Xevo TQ-S = (700011854)
ADDITIONAL INFORMATION
Before replacing parts try bringing the Treg file as described in this article: No communication with Xevo after Windows 10 upgrade KCS113014
id30889, comms, PuTTY, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD