With ACQUITY QDa Detector: dropouts in chromatographic peaks - WKB30802
Article number: 30802
SYMPTOMS
- Unstable chromatographic peaks (response sometimes drops to baseline; jagged peaks)
ENVIRONMENT
- ACQUITY QDa Detector
- Empower
CAUSE
The MS instrument method sampling rate is too high to allow appropriate chromatographic peak characterization.
FIX or WORKAROUND
- Set the sampling rate appropriately for the experiment (typically, lower it) to ensure that enough points per peak can be sampled. See the Actual vs. Intended sampling rate feedback in the MS instrument method for guidance.
ADDITIONAL INFORMATION
