Charging on Xevo TQ MS in negative polarity - WKB30566
Article number: 30566
SYMPTOMS
- Negative sensitivity constantly dropping
- Sensitivity recovers when switching to positive and then back to negative
- Even after cleaning the source components as per maintenance guide, sensitivity in negative polarity still decreases after long runs
ENVIRONMENT
- Xevo TQ MS
CAUSE
Charging on the hexapole assembly
FIX or WORKAROUND
Replace hexapole assembly (700004200)
ADDITIONAL INFORMATION