Electrospray capillary protrudes too far out of probe tip, even with adjustment rotated counterclockwise as far as possible - WKB30285
Article number: 30285
SYMPTOMS
- ESI capillary protrudes too far out of the probe tip
- Adjustment rotated all the way counterclockwise
- Capillary does not move
ENVIRONMENT
- Xevo TQ-XS
CAUSE
The API gas is not turned on.
FIX or WORKAROUND
- Turn on API gas.
- Adjust the ESI capillary downward to the appropriate position.
ADDITIONAL INFORMATION
id30285, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD