Error on ACQUITY SQD2: Source Pressure Exhaust Warning/Trip - WKB29632

Article number: 29632

SYMPTOMS

  • Error message on the console "Source Pressure Exhaust Warning / Trip"
  • The MS will not go into Operate mode
  • Tune page readbacks show a positive value, even with desolvation gas flow set to zero and API nitrogen gas switched off

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQD
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo TQ Absolute
  • Xevo TQ Absolute XR 

CAUSE

Faulty source pressure sensor

FIX or WORKAROUND

Replace the source pressure sensor (700004196 - PRESSURE SENSOR PCB ASSY).   