Low sensitivity due to mass scale and resolution drift - WKB292211

Article number: 292211

SYMPTOMS

  • Low sensitivity

ENVIRONMENT

  • Xevo TQ-XS
  • Xevo TQ-S
  • Xevo TQD
  • ACQUITY TQD
  • ACQUITY SQD
  • Xevo TQ-S micro

CAUSE

MS1 or MS2 resolution is over-resolved.

FIX or WORKAROUND

  1. Adjust resolution so that all peaks are between 0.4 and 0.6 FWHH for setup solution.
  2. Recalibrate the system.
  3. Verify sensitivity.