Severe ringing on QTof detector - WKB29199
Article number: 29199
SYMPTOMS
- Severe ringing (>50% of main peak) in one or more modes
- Poor resolution
ENVIRONMENT
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
CAUSE
The detector voltage is set too high.
FIX or WORKAROUND
- Reduce the detector voltage until the ringing disappears.
- Run detector setup.