QDa "calibration expired" error - WKB29170

Article number: 29170

SYMPTOMS

  • Calibration Expired error

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Calibration is expired.  The calibration has a defined lifetime before it needs to be repeated.

FIX or WORKAROUND

  1. Click the Resolve button in the Console
  2. Alternatively, the calibration can be run by selecting Maintain -> Check Calibration

ADDITIONAL INFORMATION

The reminder for the expired calibration can be turned off in the Health Monitor:

  1. Open Console > QDa > Configure>Health Monitor
  2. Deselect the box for Calibration Expired
    1. Note: only do this if Customer does not need the reminder, but it is not recommended to run the QDA with an expired calibration

Not able to find a solution? Click here to request help.