Lockspray crosstalk on Xevo G2-XS - WKB28396
Article number: 28396
SYMPTOMS
- Leucine spikes in the analyte function immediately after the Lockspray baffle switches from Lockspray position to Analyte position
ENVIRONMENT
- LockSpray Universal Source
- Xevo G2-XS
CAUSE
Incorrect settings
FIX or WORKAROUND
- Set Pre-filter to 2V.
- Set Ion Energy to 0.2V.
- Set Diff App 2 to 0V.
ADDITIONAL INFORMATION
id28396, XEVOG2XSTF