Lockspray crosstalk on Xevo G2-XS - WKB28396

Article number: 28396

SYMPTOMS

  • Leucine spikes in the analyte function immediately after the Lockspray baffle switches from Lockspray position to Analyte position

ENVIRONMENT

  • LockSpray Universal Source
  • Xevo G2-XS

CAUSE

Incorrect settings

FIX or WORKAROUND

  1. Set Pre-filter to 2V.
  2. Set Ion Energy to 0.2V.
  3. Set Diff App 2 to 0V.

ADDITIONAL INFORMATION

id28396, XEVOG2XSTF