Xevo G3 fails instrument setup - WKB283342
Article number: 283342
SYMPTOMS
- Instrument setup failure
- Instrument health check cards show instrument setup failure and detector setup failure
- Purging the lockmass solution multiple times does not help
- Service mode shows a Detector Setup error with the log message "Isotope Ratio Setup could not achieve IPP; beam too strong"
ENVIRONMENT
- Xevo G3
- waters_connect
CAUSE
Wrong lockmass concentration
FIX or WORKAROUND
Make new lockmass solution at a concentration of 200 pg/µL leucine enkephalin.
ADDITIONAL INFORMATION
id283342, SUPWC, XEVOG3QTOF