Signal drops rapidly in MSMS mode with Xevo G2 Tof - WKB28046
Article number: 28046
SYMPTOMS
- Signal decreases rapidly with each new injection in MSMS mode acquisitions
- Signal drop also seen during infusion in MSMS mode; signal can be recovered by switching to negative polarity for five minutes and then back to positive
- MS mode is very stable; no charging observed
ENVIRONMENT
- Xevo G2 Tof instruments
- MassLynx
CAUSE
Contamination or ion burn on the quadrupole
FIX or WORKAROUND
Remove contamination on the quadrupole using lapping film.