Skip to main content
Waters

How to replace ESI Probe Xevo Series - WKB280449

Article number: 280449

OBJECTIVE or GOAL

How to replace ESI probe

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ MS
  • Xevo TQ-S cronos
  • SQD2

PROCEDURE

Video Below

  1. Put the MS Instrument in Standby and close the API gas.
  2. Disconnect the ESI Probe inlet tubing.
  3. Disconnect the ESI Probe connector.
  4. Unscrew the ESI probe ring. And pull up the probe.                                                                                                                                                                                                                                  clipboard_ebaaa311ccb17b8995b621d470da37a45.png
  5. To install new probe, or reinstall the same Probe after service do the reverse operations, and than verify the ESI probe capillary position as described in the MS operator's guide.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.