Fluidics error during purging - WKB275284

Article number: 275284

SYMPTOMS

  • Console and tune page show error "Fluidics Stopped–Select Purge to Recover" for the sample fluidics pump
  • Error occurs randomly during purging

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo G3

CAUSE

Loose selector valve—valve pod is moving a little during switching of positions.

FIX or WORKAROUND

Tighten the selector valve retaining nut.

ADDITIONAL INFORMATION