Skip to main content
Waters

No signal on Xevo TQ series mass spectrometer - WKB271729

Article number: 271729

SYMPTOMS

  • No signal
  • Isolation valve is closed and missing sample cone

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ Absolute

CAUSE

Closed the isolation valve and removed the sample cone for cleaning but forgot to reinstall it after the cleaning.

FIX or WORKAROUND

  1. Reinstall the sample cone.
  2. Open the isolation valve.
  3. Check MSTune for the signal.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.