Skip to main content
Waters

No ETD fragmentation takes place - WKB26914

Article number: 26914

SYMPTOMS

  • No ETD fragmentation
  • Glow discharge signal is at expected intensity (mid-high e6)

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • ETD

CAUSE

Trap TriWave cell is dirty.

FIX or WORKAROUND

  1. Clean or replace the trap cell. 

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.