No flow is seen from MS sample fluidics - WKB26052

Article number: 26052

SYMPTOMS

  • See signal when reference fluidics are running
  • See no signal when trying to flow from sample fluidics

ENVIRONMENT

  • SYNAPT G2
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo TQ-S Micro
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2-S
  • SYNAPT G2-Si

CAUSE

Loose tubing

FIX or WORKAROUND

  1. Re-tighten the fittings for the sample fluidics connections.