LC line of the diverter port valve keeps popping out; no probe tip spray is seen on the Xevo TQ-S micro - WKB260994
Article number: 260994
SYMPTOMS
- Line connected to the LC port of the diverter valve on the Xevo TQ-S micro keeps popping out of the fitting
- No spray observed at probe tip
- Tightening fitting does not resolve issue
- Replacing fitting and tubing does not resolve
- If the LC line is connected directly to the top of the probe, spray from probe tip appears as expected
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Blocked port in diverter valve
FIX or WORKAROUND
- Disconnect the lines to the LC and MS ports on the Xevo TQ-S micro diverter valve.
- Connect the LC line to the MS port.
- Start the flow to backflush the port.
ADDITIONAL INFORMATION
id260994, XEVOTQSMIC