Skip to main content
Waters

Retention time shift and leak at MS diverter valve - WKB259570

Article number: 259570

SYMPTOMS

  • Weak or no signal
  • Shift in retention time
  • Leak at probe connection to MS diverter valve

ENVIRONMENT

  • MassLynx
  • Xevo TQ MS

CAUSE

Broken metal connection at the finger-tight fitting of the tool-free probe causes a leak at the diverter valve.

FIX or WORKAROUND

Replace the Tool-Free Probe (part 700011241).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.