Low response for one or more analytes in an LC-MS application - WKB260060

Article number: 260060

SYMPTOMS

  • Low response for one or more analytes in an LC-MS application
  • Sudden change in sensitivity observed
  • Sensitivity drops below usable level
  • System suitability tests fail
  • No problems found with LC or MS hardware
  • Change happens after making new mobile phase

ENVIRONMENT

  • LC-MS systems

CAUSE

Ion suppression caused by impure mobile phase

FIX or WORKAROUND

Make up new mobile phase using a different batch of solvent.

ADDITIONAL INFORMATION

This was reported after changing methanol, but it could equally be any solvent.

Because the effect is dependent on the chemical interaction of the analyte and mobile phase, not all compounds are affected by the same amount. Sensitivity could even improve for some analytes.

 

 