SYNAPT XS source turbo error after cleaning the sample cone - WKB259300

Article number: 259300

SYMPTOMS

  • SYNAPT does not pump down, gives source turbo error after cleaning the sample cone

ENVIRONMENT

  • SYNAPT XS

CAUSE

Misaligned sample cone O-ring

FIX or WORKAROUND

Remove the sample cone and confirm that the O-ring is in place.

  • If a leak around the O-ring is suspected, set the source temperature to 150 °C for one hour to improve the O-ring seating.

ADDITIONAL INFORMATION

 

id259300, SYNAPTXS