Cone gas setting is recorded incorrectly in MassLynx experimental record - WKB2591
Article number: 2591
SYMPTOMS
- Wrong cone gas value in experimental record for Xevo TQ-S and Xevo TQ MS
- Cone gas reported as OFF when it was on
- In one case, gas set to 20 L/h was recorded as off but gas flow set to 21 L/h was reported correctly
ENVIRONMENT
- MassLynx 4.2 SCN 986
- MassLynx 4.1 SCN 843
- MassLynx 4.1 SCN 876 + SCN 917
- MassLynx 4.1 SCN 810
- Xevo TQ-S
- Xevo TQ MS
CAUSE
Bug in software
FIX or WORKAROUND
Xevo TQ-S - Upgrade MassLynx software to a later version. See this article for details:
What is the latest version of MassLynx required for a Xevo TQ-S? - WKB5508
Xevo TQ MS - MassLynx 4.1 SCN 905 and SCN 950 resolve this problem,
but MassLynx 4.2 SCN 986 still shows the issue, so there is no fix for MassLynx 4.2.