Low sensitivity with UniSpray source - WKB258019

Article number: 258019

SYMPTOMS

  • Low sensitivity with UniSpray source in Unispray mode
  • Switching to ESI mode recovers the sensitivity

ENVIRONMENT

  • UniSpray source
  • Xevo TQ-XS

CAUSE

The tune page parameters are not correct. ESI tune page settings were used by mistake.

FIX or WORKAROUND

Verify and correct the tune page parameters.