No flow seen from sample sprayer on SYNAPT G2-Si - WKB25419

Article number: 25419

SYMPTOMS

  • No spray seen from sample sprayer
  • No flow elutes out of PEEK tubing that delivers to top of the ESI probe from port 3 of the Sample Select Valve
  • Plenty of fluidics wash solution and vial solution available
  • Purging does nothing to address issue
  • Port 7 of the Sample Select Valve is loose

ENVIRONMENT

  • SYNAPT G2-Si

CAUSE

Missing ferrule on fitting at port 7 of the Sample Select Valve

FIX or WORKAROUND

Install the flangeless ferrule, 1/16 inch (part 700005488).

ADDITIONAL INFORMATION

Find more details on this connection in this article: 10578.

Not able to find a solution? Click here to request help.