How to calculate Veff in MassLynx - WKB252589

Article number: 252589

OBJECTIVE or GOAL

Calculate Veff in MassLynx. 

ENVIRONMENT

  • MassLynx 4.1
  • MassLynx 4.2
  • Xevo G2-XS
  • Xevo G2-S
  • Xevo G2

PROCEDURE

  1. Open the Tune page.
  2. Select Setup > System View.clipboard_e1262348fbd36c27d80ead83b0cb748fa.png
  3. Enter password: access
  4. Select System > Acquisition Settings > Nominal Mass.
  5. Select the check box for "Calculate and Set Veff".
  6. Reference Mass = molecular mass of Leu Enk (556.27 Pos mode) Measured Mass = Actual Mass of Leu Enk
  7. Click Calculate > Update > Close.clipboard_e358d326f8c4e3f7040d0c81b4aed00bf.png
  8. Verify that the new mass position is correct. 
  9. Repeat if necessary.