Calibration invalid for the analysis type when trying to run Sample Tune and Develop Method from IntelliStart - WKB252344
Article number: 252344
SYMPTOMS
- Calibration invalid for analysis type when trying to run Sample Tune and Develop Method from IntelliStart
- Running calibration from IntelliStart to create a new calibration file fails
- Previous calibration file exists and is linked to the current tune page
- Data acquired with previous calibration file was acceptable
ENVIRONMENT
- MassLynx
- Xevo TQ-S micro
CAUSE
Unknown
FIX or WORKAROUND
- In the MS Console, highlight IntelliStart under the Xevo TQ-S micro in the system list on the left-hand side.
- From the Configure menu, select IntelliStart Configuration.
- Clear Invalid Calibration.
- Save changes.
- On the Tune page, go to Calibrate > Link Calibration.
- Select the existing calibration file that was known to work.
ADDITIONAL INFORMATION