The Edwards RV5 roughing pump on the QDa has oil bubbling around the flange that connects the oil mist filter - WKB243878

Article number: 243878

SYMPTOMS

  • Oil bubbles around the connection of the exhaust at the oil mist filter
  • Excess oil buildup on the top of the pump

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

The flange is not tightened.

FIX or WORKAROUND

  1. The flange where the oil mist filter connects is flat on two sides.
  2. Use an adjustable wrench to tighten the flange . 

 