Skip to main content
Waters

Instrument setup struggles to pass negative ion autotune on ACQUITY RDa or BioAccord - WKB242805

Article number: 242805

SYMPTOMS

  • Instrument setup fails for negative ion mode
  • Autotune fails for negative ion mode
  • The backing was adjusted to bring it into the appropriate range (per wkb203528)
  • PM was performed
  • There is a tail on the peak that does not tune out
    Seth tail demo lab 3.PNG

ENVIRONMENT

  • ACQUITY RDa
  • BioAccord
  • Instrument setup
  • Autotune

CAUSE

Misalignment of the front end, a leak, or both

FIX or WORKAROUND

  1. Replace:
    1. The gasket with stent tube that is behind the ion block - 700009603
    2. Cone clamp/PEEK cover on the ion block for the cone - 700009626
    3. The pumping block - 700012789
  2. Pump down. 
  3. Place the instrument in Operate with gas on for an hour.
  4. Run instrument setup.

ADDITIONAL INFORMATION

 

id242805, RDA, SUPWC

Not able to find a solution? Click here to request help.