"Nitrogen Gas Failure" error appears in MassLynx and "Reset Nitrogen Failure" is dimmed/unavailable on the MS Tune page - WKB242392

Article number: 242392

SYMPTOMS

  • After an N2 failure, the N2 is restored
  • The "Nitrogen Gas Failure" error still shows, but "Reset Nitrogen Failure" is grayed out on the tune page

ENVIRONMENT

  • MassLynx
  • Xevo TQ-S micro
  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Unknown

FIX or WORKAROUND

  1. Open the MS Console.
  2. Click on the MS.
  3. Under Control, reset the MS (for example, "Reset Xevo TQD").

ADDITIONAL INFORMATION

If doing a Control > Reset MS  in the MS console does not resolve the issue, resetting the electronics may be required. 