"Nitrogen Gas Failure" error appears in MassLynx and "Reset Nitrogen Failure" is dimmed/unavailable on the MS Tune page - WKB242392
Article number: 242392
SYMPTOMS
- After an N2 failure, the N2 is restored
- The "Nitrogen Gas Failure" error still shows, but "Reset Nitrogen Failure" is grayed out on the tune page
ENVIRONMENT
- MassLynx
- Xevo TQ-S micro
- SQ Detector 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Unknown
FIX or WORKAROUND
- Open the MS Console.
- Click on the MS.
- Under Control, reset the MS (for example, "Reset Xevo TQD").
ADDITIONAL INFORMATION