Turbo speed only reaches 70% - WKB2421

Article number: 2421

SYMPTOMS

  • Turbo speed only reaches 70% then times out causing roughing pump to stop

ENVIRONMENT

  • SQD, 3100, SQD2, TQD
  • Xevo TQD, Xevo TQ, Xevo  TQ-S, Xevo TQ-S micro, Xevo TQ-XS
  • ZQ, Quattro Micro, Quattro Ultima, Quattro Premier / XE
  • Xevo G2

CAUSE

Leak in ion block 

FIX or WORKAROUND

  1. Replace all o rings in the ion block

ADDITIONAL INFORMATION