Tubing pops off the diverter valve on TQ detector at normal flow rate - WKB24176
Article number: 24176
SYMPTOMS
- Tubing pops off the diverter valve or the top of the ESI probe
- Replaced tubing and fittings from the column to the diverter valve
- Replaced tubing and fittings from the diverter valve to the ESI probe
- Confirmed flow through the diverter valve
ENVIRONMENT
- TQ Detector
- SQ Detector
- 3100 Mass Detector
- MassLynx
CAUSE
Blocked ESI probe
FIX or WORKAROUND
Rebuild the ESI probe using the ESI Probe Service Kit (700005795).
ADDITIONAL INFORMATION