Tubing pops off the diverter valve on TQ detector at normal flow rate - WKB24176

Article number: 24176

SYMPTOMS

  • Tubing pops off the diverter valve or the top of the ESI probe
  • Replaced tubing and fittings from the column to the diverter valve
  • Replaced tubing and fittings from the diverter valve to the ESI probe
  • Confirmed flow through the diverter valve

ENVIRONMENT

  • TQ Detector
  • SQ Detector
  • 3100 Mass Detector
  • MassLynx

CAUSE

Blocked ESI probe

FIX or WORKAROUND

Rebuild the ESI probe using the ESI Probe Service Kit (700005795).

ADDITIONAL INFORMATION

 

 

id24176, A-SQ, Electrospray, Q3100MSD, SQD, TQD, TQDIVD

 