Skip to main content
Waters

Vion instrument setup fails (Resolution setup in Quartz) - WKB241723

Article number: 241723

SYMPTOMS

  • Resolution setup fails in Quartz
  • 1e7 LeuEnk response in Pos Sens

ENVIRONMENT

  • Vion

CAUSE

The wrong protocol was followed, and the LeuEnk concentration ended up higher than required.

FIX or WORKAROUND

Make up LeuEnk at the correct concentration, purge fluidics, and run the setup again.

ADDITIONAL INFORMATION

 

id241723, VION

Not able to find a solution? Click here to request help.