Turbo controller fault on the Xevo TQ-S micro - WKB240093

Article number: 240093

SYMPTOMS

  • Turbo controller fault
  • High exhaust pressure trip
  • Instrument vents during a run
  • Able to pump down without error

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQ-S cronos
  • Xevo TQ-S
  • Xevo TQ-XS

CAUSE

The exhaust tubing from the backing pump is crimped.

FIX or WORKAROUND

Replace the tubing and make sure that it has no sharp bends.

ADDITIONAL INFORMATION