Vion drops out of IMS pressure lock after cone cleaning - WKB238028
Article number: 238028
SYMPTOMS
- After the regular cone cleaning, the instrument often drops out of IMS pressure lock
ENVIRONMENT
- Vion IMS QTof
CAUSE
The cone O-ring does not seat correctly all of the time.
This causes variable backing pressure (and therefore trap pressure) such that it fails its 20% tolerance.
FIX or WORKAROUND
- Note the backing pressure before removing the cone.
- Ensure that the cone is seated correctly when replacing, by ensuring that the pressure matches the previous value.
ADDITIONAL INFORMATION
id238028, VION